UTRED-16 – USB PDF EXTERNAL PROBE TEMPERATURE LOGGER

  • Ideal for multi-stage distribution channels involving bulk through to “last-mile” stages.
  • Display showing temperature and alarm status of up to six alarms.
  • Cost effective multi-use format with a two year battery life.
  • Automatic generation of a PDF report as a permanent record of the data.

The LogTag® UTRED-16 is compatible with all external probes from the ST100 range, and the extension cable.

The display shows recording status, current temperature, alarm status of up to six user-configurable alarms and duration of any temperature excursions.

When the UTRED-16 is connected to a computer using the integrated USB plug, a detailed PDF report of the recorded temperature data and alarms can be automatically generated without the need for extra accessories and proprietary software.

For more detailed data analysis, the UTRED-16 also creates an encrypted data file compatible with LogTag Analyzer® operating software and data management options.

SPECIFICATIONS

Product ModelUTRED-16F (Fixed Battery).
Sensor Measurement Range-40°C to +99°C (-40°F to +210°F).
Operating Temperature Range-30°C to +60°C (-22°F to +140°F).
Storage Temperature Range0°C to +40°C (32°F to +104°F).
Rated Temperature Reading Accuracy Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F), typically ±0.3°C (0.6°F).
Better than ±0.8°C (±1.4°F) for -40°C to -20°C (-40°F to -4°F), typically ±0.5°C (0.9°F).
Better than ±0.8°C (±1.4°F) for +40°C to +70°C (+104°F to +158°F), typically ±0.7°C (1.3°F).
Better than ±1.2°C (±2.2°F) for +70°C to +99°C (+158°F to +210°F), typically ±1°C (1.8°F).
Rated Temperature Reading Resolution Better than 0.1°C (0.2°F) for -30°C to +60°C (-22°F to +140°F).
Better than 0.2°C (0.4°F) for -40°C to -30°C (-40°F to -22°F).
Better than 0.5°C (0.9°F) for +60°C to +99°C (+140°F to +210°F).
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
Sampling IntervalConfigurable from 30 seconds to 18 hours.
Logging Start OptionsPush button start or specific date & time. Optional start delay of up to 72 hours.
Recording Indication“REC” symbol on LCD.
Download Time Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report.
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured).
EnvironmentalIP61 (when vertically mounted or hung with USB dust-cap in place).
Power SourceCR2450 3V LiMnO2 (Fixed battery).
Battery Life1 year storage, 2 years of normal use (based on 15 minute logging, download data monthly).
Real Time Clock Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Connection InterfaceUSB 2.0, A-type plug.
SoftwarePDF Reader, LogTag® Analyzer 3.1.9 or higher.
PDF ReportOne page report with trip and alarm summary. Multi-page report with list of readings, including date and time.
Size93mm(H) x 54.5mm(W) x 8.6mm(T) including protective USB cap.
Weight39g.
Case MaterialPolycarbonate.

Compliance and Certifications
FDA CFR21 Part 11 FCC EMC Directive CE EMC Directives Conforms to RoHS