Products


LogTag® TRID30-7 :
Temperature Recorder with 30day summary display

The LogTag® TRID30-7 temperature recorder features a display together with a data logging function storing up to 7770 temperature readings.  Statistical temperature and duration readings for up to 30 days can be reviewed on the display.

The visual display of current temperature and previous alarms is an important feature in “static” applications such as cool rooms and refrigerators. The TRID30-7 can also be used in transit applications where a direct readout of the statistical data is required without use of a computer or readout device.

The display arrangement is designed to show ‘at a glance’ if temperature violations have occurred during the current day and up to the previous 29 days. The display also shows the current temperature reading, the current time, recording status and battery status.

If a reading outside the pre-set “Alarm” limits is recorded at any time, a “day alarm indicator” appears on the display. A visual alarm trigger is displayed if one or more of the configured alarm trigger conditions have been met. An alarm trigger condition can be a single reading above the upper or below the lower threshold values, a set of consecutive violating readings or an accumulation of violation readings encountered.

Details of any violations can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged data. Logged data can be downloaded via a standard LogTag Interface to the free companion software LogTag Analyzer which provides facilities for displaying data in chart, table or statistical formats and allows electronic archiving, export or transmission of the data in support of sophisticated data management systems.

The recorder incorporates an external temperature sensor uniquely arranged in a recess in the case.
This arrangement protects the sensor from physical damage while retaining the faster response time normally associated with remote sensors.

The recorder can be configured to suspend processing of readings for alarms and max/min statistics for a period of time after button press activity. This allows the user to review the current statistics or clear an alarm without causing a false alarm or statistic while handling the recorder.

Two different battery options are available:-
Part# TRID30-7R : Equipped with user replaceable battery (CR2032) with typical battery life of 1 year.
Part# TRID30-7F : Fixed cell (non user replaceable) with typical battery life of 2-3 years.

 
Highlights
Data Logging – Records time and temperature data for later download
Day alarm summary - Displays in calendar-like format alarm triggers for up to the last 30days
Current time & temperature - Shows the current time and temperature when not in review mode
Push Button Logging start with optional delayed start or specific Time & Date start
Rapid Download! Takes only seconds to download recordings
‘Pre-Start’ logging - Logtag can be configured to record even if it has not been started.
High performance at low cost
Thin flat case – Fits easily into packaging and thin enough to be easily mailed “letter rate”.
Real time clock records time & temperature simultaneously
Easy to use LogTag Analyzer software that runs on any PC configures LogTag for recording then downloads resulting data for analysis. Data can also be exported to fomats compatible with other applications such as Excel
Re-calibration to achieve higher accuracy possible
   

Product Brochure

 
QuickStart Guide
 
Product Specification & Guide
 

Specifications
Model#s
 
TRID30-7R : (replaceable battery, download instructions here)
TRID30-7F : (fixed battery)
TRID30-7FW (fixed battery, preconfigured and package to meet WHO specification E006/TR06.3)
Measurement Range
 
-30°C ~ +60°C (-22°F ~ +140°F)
Display Resolution
 
0.1°C(0.2°F) for measurements -30°C ~ +40°C (-22°F ~ +104°F)
0.2°C(0.4°F) for measurements above +40°C (+104°F)
Accuracy
 
Better than ±0.5°C (±0.9°F) for measurements from -20°C~+40°C (-22°F~+104°F) -typically ±0.3°C (0.6°F)
Better than ±0.8°C (±1.5°F) for other measurements - typically ±0.5°C (0.9°F).
Sensor reaction time
 
typically less than 5 minutes(T90) in moving air (1m/s).
Capacity
 
Data logging memory : 7770 logs ( 53 days @ 10min logging, 80 days @ 15min logging)
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min and duration values
Sampling frequency
 
adjustable, 30 seconds to several hours
Logging start options
 
Push button start or specific date & time. Optional start delay of up to 18hours.
Recording indication
 
State indicator "RECORDING"
Download Time
 
Typically with full memory (7770 readings) in less than 5 secs depending on computer or readout device used.
Environmental
 
IP65 (roughly equivalent to NEMA 4) when vertically mounted or hung.
Power source
 
3V Lithium-Manganese Dioxide extended temperature chemistry
Battery life
 
TRID30-7R : typically 1 year of operation
TRID30-7F : typically 2+ years of operation
NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more that once daily for no longer than 30 seconds each time. Download data monthly.
Size
 
93.0mm(H)x54.5mm(W)x8.6mm(T)
Weight
 
TRID30-7R : 39grams
TRID30-7F : 43grams
Case Material
 
Polycarbonate
Other features

 

• Optional clearing of alarm indication by push button (places inspection mark at same time).
• Download inspection mark recorded in LogTag memory every time LogTag is downloaded.
• 'Prestart' fail-safe logging (records temperature data even if not started)
• Low battery indication.
• Comprehensive Alarm trigger configuration.
• Calibration to achieve higher accuracy possible.
•Compatible with standard LogTag Interface cradle

Note: Resolution & Accuracy are rated values - actual product performance may be substantially better than figures stated.

Standards compliance
EMC Compliance

Tested and complies with EC EMC directives (EN 61000-6-3:2006 & EN 61000-6-1:2005)
see LogTag TRID30 CE mark EMC declaration of conformance

Tested and complies FCC Part 15 Subparts A and B as an unintentional radiator..
see LogTag TRID30 FCC EMC declaration of conformance

WHO Compliance
PQS

FDA CFR21 Part 11 Compliance
CFR21Part11

Designed for FDA CFR21 Part 11 Digital signatures
 
Manufactured to meet the European RoHS Directive
The RoHS Directive stands for "the restriction of the use of certain hazardous substances in electrical and electronic equipment". This Directive bans the placing on the EU market of new electrical and electronic equipment containing more than agreed levels of lead, cadmium, mercury, hexavalent chromium, polybrominated biphenyl (PBB) and polybrominated diphenyl ether (PBDE) flame retardants.

 

Accessories

 


INTERFACE CRADLES

Logtag's unique interface cradle design provides rapid and reliable LogTag data transfer to Logtag Analyzer software in a matter of seconds. By far the fastest in it's class.
Interface cradles are available in USB and RS232 connection versions.

Click here for Interface Cradle information area

 
 


Wall Mount Bracket

The LogTag Wall Mount Bracket provides an easy and tidy solution to mounting a LogTag on a wall,door or vertical surface while still allowing easy removal of the LogTag for inspection or download.

Click here for Wall Mount Bracket details

 
 

 

Protective Enclosure

The LogTag® Protective Enclosure is specifically designed for use with LogTag recorder products and allows LogTags to be used in harsh environments.

Click here for Protective Enclosure details


 

 


Last updated May 9, 2013 , © Copyright LogTag Recorders Limited