SRIL-8 SINGLE-USE LOW TEMPERATURE LOGGER

  • Records temperature from +40°C to as low as -80°C.
  • Single-trip dry ice Temperature Logger.
  • 8000 readings. Rated for a typical trip of 2 weeks at dry ice temperatures.
  • ‘OK’ & ‘ALERT’ Indicator with high and low customizable alarms.

The LogTag® Dry Ice “Probe-less” Temperature Logger operates, measures and stores up to 8000 temperature readings in temperature environments ranging from -80°C to +40°C (-112°F to +104°F).

Intended for use in transit monitoring of articles stored in packaging incorporating dry ice cooling agents.

Available in multi-trip (TRIL-8) version.

Using the LogTag® Interface and LogTag’s freely available companion software LogTag Analyzer, the LogTag® Dry Ice Temperature Recorder is easily configured for recording conditions including delayed start, sampling interval, number of readings and configuration of conditions to activate the ALERT indicator.

The Red Alert indicator provides an immediate indication, without access to a PC, if any readings are outside the limits specified at the time the unit was configured.

Green OK indicator provides immediate visual confirmation, without access to a PC, that the unit is operating.

Readings are downloaded using LogTag Analyzer which provides facilities for charting, zooming, listing data statistics and allows exporting the data to other applications such as Excel.

SPECIFICATIONS

Product ModelSRIL-8.
Sensor Measurement Range-80°C to +40°C (-112°F to +104°F).
Operating Temperature Range-80°C to +40°C (-112°F to +104°F).
Storage Temperature Range-20°C to +40°C (-4°F to +104°F).
Rated Temperature Reading Accuracy Better than ±1°C for -30°C to +20°C.
Better than ±1.2°C for -45°C to -30°C and +20°C to +40°C.
Better than ±1.7°C for -80°C to -45°C.
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration.
Rated Temperature Reading Resolution Less than 0.1°C for -80°C to 0°C.
Less than 0.2°C for 0°C to +20°C.
Less than 0.5°C for +20°C to +40°C.
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
Sensor Reaction TimeTypically less than 5 minutes (T90) in moving air (1m/s).
Recording Capacity8000 temperature readings. 53 days @ 10min logging, 80 days @ 15min logging.
Sampling IntervalConfigurable from 1 minute to 18 hours.
Logging Start OptionsPush button start or specific date & time.
Recording IndicationFlashing ‘OK’ indicator / flashing ‘ALERT’ indicator.
Download TimeTypically less than 5 seconds for full memory (8000 readings), depending on computer or readout device used.
EnvironmentalIP65 (roughly equivalent to NEMA 4).
Power Source3.6V low temperature chemistry Lithium battery.
Battery LifeStorage life of up to 12 months before ‘start’. Rated for a typical trip of 2 weeks duration at dry ice temperatures.
Real Time Clock Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Size86mm(H) x 54.5mm(W) x 8.6mm(T).
Weight35g.
Case MaterialPolycarbonate.

Compliance and Certifications
FDA CFR21 Part 11 FCC EMC Directive CE EMC Directives Conforms to RoHS

View other LogTag® products